A&D EJ-1500 Scale User Manual


 
33
2. CC 0 is displayed.
3. If necessary to change the weight value to check, press the SAMPLE key and
change the weight value using the following keys.
SAMPLE key To select the digit blinking to change.
RE-ZERO key To set the value of the digit selected.
PRINT key To store the value and return to step 2.
4. At step 2, press the PRINT key. The zero point
is weighed and the weighed value is displayed for a
few seconds.
5. Place the displayed weight on the pan and press
the
PRINT key to weigh it. The weighed value is
displayed for a few seconds.
6. end appears.
7. glp is displayed and calibration test report is
output.
8. end appears again. Remove the weight and
press the MODE key to return to the weighing
mode.
CC 0
0.0
CC 0
The weighed data
is shown with “g”.
With nothing
on the pan
2000.0
2000.0
2000.0
Displayed
weight
end
end
The weighed data
is shown with “g”.
Returns to
the weighing mode
glp
GLP output