433201A V1.5 05/07
4
Operation
Measurement Mode
1. Open the meter lid and press the POWER/MODE button to turn the meter on.
“MEASURE” will be displayed on the LCD screen. Check the battery if the meter
display does not switch on.
Note: For best accuracy, remove any input connectors before turning the unit
ON. The meter will not perform a correct self calibration if there is an input
connected.
2. Select the thermocouple type using the TYPE button. The T/C type selected will be
indicated along the bottom of the display. Each time the TYPE button is pressed
the indicator will advance one type.
3. Select
o
C or
o
F via the
o
C/
o
F button.
4. Carefully insert the thermocouple probe into the sub-miniature female connector on
the side panel.
5. Read the temperature on the LCD display.
6. To turn the meter off, press and hold the POWER/MODE button for more than 2
seconds.
Offset Adjustment
Thermocouples produce an output signal that is non-linear. In some instances it may be
necessary to offset the reading to compensate for the non-linearity of the thermocouple.
1. In Measurement Mode, press and hold the /buttons together until ”OFS” is
displayed on the LCD.
2. Use the / buttons to adjust the offset from -9.0 to 9.0
o
F (-5.0 to 5.0
o
C)
3. Press and hold the /buttons together for less than 2 seconds to save the offset
and return to Measurement Mode.