A SERVICE OF

logo

42545A-en-GB_V4.4 12/14
7
Specifications
Infrared Thermometer Specifications
Range / Resolution -50 to 1000
o
C (-58 to 1832
o
F) 0.1
o
C/F
Accuracy
(of reading)
± (2% of reading + 9°F/4°C) <30°F (-1°C)
± (2% of reading + 4°F/2°C) 30°F to 800°F (-1°C to 426°C)
± (2.5% of reading + 6°F/3°C) 800 to 1000°F(426 to 537°C)
± (3% of reading + 9°F/4°C) >1000°F/537°C
Note: Accuracy is specified for the following ambient temperature range:
18 to 28°C (64 to 82°F)
Emissivity Adjustable from 0.1 to 1.00 (0.95 default value)
Field of View D/S = Approx. 50:1 ratio (D = distance, S = spot)
Laser power Less than 1mW (Class II)
Spectral response
8 to 14 m (wavelength)
General Specifications
Display digit backlit LCD display with function indicators
Display rate 1 second approx.
Operating Temperature 0°C to 50°C (32°F to 122°F)
Operating Humidity Max. 90% RH
Power Supply 9V battery
Automatic Power Off Approx. 7 seconds after the trigger is released
Safety compliance CE
Weight 290g / 10.2 oz
Dimensions 100 x 56 x 230mm (3.9 x 2.2 x 9.0”)
Copyright©2014FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com