9
cleaning &maintenance
Collection Bole:
1. Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the
DC input receptacle on the unit.
2. With the collection bole still in the holder, remove the lid. The bole can now be taken out of
the holder to be emptied.
NOTE
Collection bole should be emptied and cleaned aer each use. The 800ml disposable collection
bole and lid are meant for single-patient use
3. Collection boles and lids should be thoroughly cleaned aer each use by one of the following
methods:
All Boles:
. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then
wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.
Autoclave for 3 to 5 minute cycle at 270
o
F (132
o
C).
In lieu of autoclave sterilization, the above parts can be soaked in an activated dialdehyde
solution.
Suction Unit:
1. With the power switch in the “OFF” position, disconnect the Drive Suction Unit from all exter-
nal power sources.
2. Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
CAUTION
Do not submerge in water as this will result in damage to the vacuum pump.
Tubing:
1. Disconnect the tubing from the unit.
2. Tubing should be rinsed thoroughly aer every use by running hot tap water through it fol-
lowed by a solution of one part vinegar to three
parts hot water.
3. Rinse with hot tap water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
Carrying Case:
1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.