![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/f/72/f7279902-05c8-486c-ad97-ca95733f3b83/f7279902-05c8-486c-ad97-ca95733f3b83-bg28.png)
DELTA LTL2000S(Q) Retroreflectometer 36
APPENDIX B
Status code interpretation
Rl Status code interpretation:
bit pattern value Description
Status Code: 00000001 1 Error in Rl Result
Status Code: 00000010 2 Stray Light Warning
Status Code: 00000100 4 Log full Warning
Status Code: 00001000 8 Not defined
Status Code: 00010000 16 Low Battery Warning
Status Code: 00100000 32 High Zero signal Warning
Status Code: 01000000 64 High Signal with Lamp on Error
Status Code: 10000000 128 High Signal with Lamp off Error
Qd Status code interpretation:
bit pattern value Description
Status Code: 00000001 1 Error in Qd Result
Status Code: 00000010 2 Stray Light Warning
Status Code: 00000100 4 Not defined
Status Code: 00001000 8 Not defined
Status Code: 00010000 16 Low Battery Warning
Status Code: 00100000 32 Not defined
Status Code: 01000000 64 High Signal with Lamp on Error
Status Code: 10000000 128 High Signal with Lamp off Error
E.g. Status code 18 is composed of Low Battery Warning (16) + Stray Light Warning (2)