A SERVICE OF

logo

Pumps & Conveying System Components Chapter 6: Troubleshooting 43
Problem
Possible Cause
Possible Remedy
Blower is dead-headed.
Check for obstruction.
Vent valve is not operating.
Check for proper signal,
voltage, and 80 to 90 psi
(552 to 621 kPa)
compressed air.
Worn seals in blower.
Rebuild/replace blower.
Worn bearings in blower.
Rebuild or replace.
Loose drive belt(s). Tighten
or replace if worn.
Follow the vacuum path from
the blower through the filter,
piping, valves and material
receivers to isolate where the
loss occurs.
Filter chamber problem. The
vacuum at the filter chamber
should be the same as the
blower generates. If the
vacuum at the filter chamber
is adequate, the problem is
down line.
Vacuum conveying rate is
declining due to a vacuum loss
in the system.
Pump package blower
problem. Use a cfm/cfh
monitoring device to ensure
that blower cfm/cfh is to the
manufacturer’s specifications.
Consult Sevice Engineer if
problem persists.
Dirty filter. Clean or replace
filter.
Loose clamps, gaskets or
couplers. Tighten or replace.
Filter chamber discharge
flapper not sealing under
vacuum. Check for proper
operation; clean or repair as
needed.
Check for faulty compressed
air blowback solenoid.
Check for voltage signal
during cleaning cycle,
proper solenoid operation,
incorrect wiring, proper
compressed air connection,
and for worn plunger in
valve.