Welch Allyn Medical Diagnostic Equipment LXi Blood Pressure Monitor User Manual


 
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA
(p) 800.535.6663 (f) 315.685.2174
www
.welchallyn.com
SureTemp and Spot Vital Signs are registered trademarks of Welch Allyn, Inc.
© 2005 Welch Allyn. All rights reserved. SM2871 Rev B
Welch Allyn Spot Vital Signs
®
LXi
Specifications
Only Masimo SET SpO
2
sensors, including LNOP durable adhesive and
NR value disposable sensors should be used with Masimo pulse
oximetry. Masimo, Masimo SET, LNOP, and NR are registered
trademarks of Masimo Corporation.
Only Nellcor
OxiMax
®
pulse oximetry sensors should be used
with the Nellcor pulse oximetry option. Nellcor and OxiMax are
registered trademarks of Nellcor Puritan Bennett, Inc.
Noninvasive Blood Pressure (NIBP)
C
uff Pressure Range 0 to 300 mmHg (0 to 40.0 kPa)
S
ystolic Range 60 to 250 mmHg (8.0 to 33.3 kPa)
D
iastolic Range 30 to 160 mmHg (4.0 to 21.3 kPa)
M
ean Arterial 40 to 190 mmHg (5.3 to 25.3 kPa)
Pressure Range
Blood Pressure Accuracy Meets or exceeds AAMI SP10:2002
Overpressure Cutoff 315mmHg ±15 mmHg (42.0 kPa ± 1.9 kPa)
Blood Pressure Typical: 15 seconds
D
etermination Time
Pulse Rate Range 35 to 199 bpm
Pulse Rate Accuracy
±5.0%
Temperature
T
emperature Range SureTemp
®
P
lus: 80° F to 110° F (26.7° C to 43.3° C)
Braun ThermoScan® PRO 4000: 68° F to 108° F (20° C to 42.2° C)
Calibration Accuracy SureTemp Plus: ±0.2° F (0.1° C) (Monitor Mode)
Braun ThermoScan PRO 4000 for displayed temperature ranges:
± 0.4° F (±0.2° C) 95.9° F to 107.6° F (35.5° C to 42° C), ±0.5° F
(±0.25° C) (outside this temperature range)
Display Resolution 0.1° F or ° C
SureTemp Plus Predict Time Oral Approx. 4 to 6 seconds
Adult Axillary Approx.12 to 15 seconds (age 18 years and older)
Pediatric Axillary Approx. 10 to 13 seconds (age 17 years and younger)
Rectal Approx. 10 to 13 seconds
Probe and Probe Well Oral/Axillary Blue
Rectal Red
SpO
2
Performance Measurement 1 to 100%
Range 25 to 240 bpm
SpO
2
Accuracy 70% to 100%
Pulse Rate Accuracy 25 to 240 bpm ±
3 digits
Mechanical
Dimensions Height 10.63 “ (27 cm)
Length/PRO 4000
8” (20.32 cm)
Length/SureTemp Plus 7.5” (19 cm)
Depth 5.25” (13.34 cm)
Weight 7.5 lbs (3.4 kg)
Electrical
AC power supply: 8V DC at 0.75 A
Rechargeable sealed lead acid internal battery
Battery recharge time: 90% to 100% after 6 hours; 1 additional hour for PRO 4000;
4 additional hours for external printer
Simultaneously operates device and charges battery when AC power adapter is plugged in
Wireless Radio
Wireless Network Interface IEEE 802.11b DSSS, WiFi compliant
Frequency 2.4 to 2.4835 GHz
Environmental
Operating Temperature 50° F to 104° F (+10° C to +40° C)
(Device)
Operating Relative 15% to 95% (noncondensing)
Humidity
Operating Altitude -557 ft to +16,000 ft (-170 m to +4,877 m)
Storage Temperature Device with SureTemp Plus: -13° F to 131°F (-25° C to 55°C)
Device with PRO 4000: -4° F to 122°F (-20° C to 50°C)
Storage Humidity 15% to 95% (noncondensing)
EMC Standard IEC/EN 60601-1-2:2001
Warranty
Two-year warranty on device parts and labor
• Service contracts available
Braun ThermoScan® is a registered trademark
of the Braun Company.