CARE AND CLEANING
1. The Remote Fever Monitor is a delicate
electronic device. To insure proper functioning,
do not immerse either unit in water and avoid
dropping them.
2. Clean both units with a soft cloth moistened
with either water or alcohol. Never use an
abrasive.
3. Keep out of the reach of children and in the
storage case when not in use. The optimal
storage temperature is between -4°~140°F/-
20°~60°C.
4. Do not disassemble the thermometer.
If you believe your Remote Fever Monitor is
not operating properly, please check the
batteries first. If it still does not operate
normally, or you are having other technical
problems, please contact:
Unisar Inc.
15 West 36th Street
New York, NY 10018
Tel. 1-888-232-6476
TROUBLESHOOTING
Poor Reception
Cause: Child’s Thermometer is not sending a
signal or the distance between both units is
beyond the specified range for over 40 seconds.
Effect: Parents’ Unit will beep 3 times at 10
second intervals and the receiving symbol will
disappear and SCAN will appear.
Solution: Check the position of the thermometer
and/or move back into the transmission range.
Low Battery
Cause: Battery power is too low to operate the
unit.
Effect: Parents’ Unit will beep every 5 minutes
and the low battery symbol will appear.
Solution: Change the battery in whichever unit
is causing the problem or in both units.
Temperature Too Low
Cause: Thermometer lost contact with baby’s
skin or the baby’s temperature is too low.
Effect: The Parents’ Unit will beep for 10 sec-
onds.
Solution: Check your child and the position of
the thermometer immediately.