Topcom 5331 WHO Blood Pressure Monitor User Manual


 
10
TOPCOM BPM Wrist 5331 WHO
To erase all data from one memory zone:
•Press the START/STOP button to activate the display.
•Press the UP or DOWN button to select the desired memory
zone.
Press the MEMORY RECALL button to see the average of all
measurement results.
•Press the MODE and the UP button simultaneously to erase
all data from the selected memory zone.
15 Technical specifications
*Specifications are subject to change without notice.
Power source
Measurement method
Measurement zone
Accuracy
Pressure sensor
Inflation
Deflation
Memory capacity
Auto-shut-off
Operation environment
Storage environment
Weight
Wrist circumference
Limited patient
Dimensions
K.B.
Classification
3V DC: Two LR03 (AAA) Non rechargeable batteries
Oscillometric
Pressure: 0 ~ 300 mmHg
Pulse: 40 ~ 199 beats/minute
Pressure: ± 3 mmHg
Pulse: within ± 5% of reading
Semi-conductor
Pump driven
Automatic Pressure release valve
3 Memory zones for each 30 memories
1 minute after last key operation
+10°C ~ + 40°C
< 85% RH
-20°C ~ + 70°C
< 85% RH
120g (without batteries)
135 ~ 195 mm
Over 18 years old
74(L) x 84(W) x 32(H) mm
Short time operation 2 minutes
Internally powered equipment
Classification of type BF (applied part)
IPX0
Not suitable for use in presence of flammable
anaesthetic mixture with air or with oxygen or
nitrous oxide
Continuous operation with short-time loading