Teledyne Analytical Instruments 37
RESET menu again. The data set can be downloaded using the LOG
TRANSMIT screen (see Section 3.2.17).
To start the data logger:
1. Use the UP/DOWN keys to navigate to the LOG RESET &
START screen and press ENTER.
2. When the screen begins to blink press ENTER again to start
the data logger. Once the data logger has started, the screen
changes RESET&START to STOP. During logging a
blinking “L” appears in the upper left corner of the screen.
To stop the data logger:
1. Press ENTER to stop the data logger and return to the LOG
RESET & START screen.
3.2.18 LOG TRANSMIT Screen
Data can be downloaded to a computer using the optional RS-232
port and cable attached to a PC. The RS-232 port accepts a standard data
cable with a DB-9 connector. The computer must be able to accept data
from a RS-232 source with the following characteristics:
9600 baud
8 bit data
no parity