Suburban Mfg 12155-00-EPEPAA Oxygen Equipment User Manual


 
32 SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Troubleshooting
There is excessive oxygen leakage from the cage.
Note: Some oxygen leakage is normal. The corners of the cage
are left open to help distribute the oxygen in the cage uniformly and
to allow for the escape of carbon dioxide. If you feel that the
amount of leakage is excessive, perform the following steps.
First: Make sure the door is properly closed and latched. If the
door does not close or latch properly, the latch brackets are
probably out of adjustment. Refer to Latch Adjustment on
Page 11.
Second: Make sure the oxygen supply pressure is not set too
high.
Third: There are black rubber seals on all four sides of the door.
Check all four door seals to make sure that none are missing, loose,
or damaged. If a seal is loose, tighten the mounting screws/nuts on
that seal. If a missing or damaged seal is found, call SSCI and order
a replacement. Refer to Page 25 for part numbers and replacement
instructions.
Remedial Action