Shop-Vac XPO100 Oxygen Equipment User Manual


 
SECTION 2—COMPONENT REPLACEMENT
XPO
2
10 Part No 1150748
1. Remove label from the top of the XPO
2
(FIGURE 2.2).
NOTE: This label adheres to the front and rear sections of the XPO
2
. This label MUST be removed
to separate the two sections.
FIGURE 2.2 Remove/Install label
2. Position the XPO
2
so that the back panel label is facing up (Detail “A” in FIGURE 2.3).
3. Remove the four mounting screws from the back of the XPO
2
using a T15 Torx
®
screwdriver (Detail “A” in FIGURE 2.3).
FIGURE 2.3 Removing the Front Cover
Label
XPO
2
Mounting Screws
Mounting Screws
Ribbon Cable
Front Cover
DETAIL “A”
DETAIL “B”
Back Panel Label
Tubing
Inlet Filter
DETAIL “C”
Front Cover
Ribbon Cable
PC Board