ResMed VPAP Auto Respiratory Product User Manual


 
ResMed Corp Poway, CA, USA +1 858 746 2400 or 1 800 424 0737 (toll free), ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). Offices in Austria, Brazil, Finland, France, Germany,
Hong Kong, Japan, Malaysia, Mexico, Netherlands, New Zealand, Singapore, Spain, Sweden, Switzerland, United Kingdom (see website for details). ResControl, ResLink, ResScan, ResTraxx, S8, TiControl, VPAP Adapt
SV and Vsync are trademarks of ResMed Ltd. AutoSet, HumidAire and VPAP are trademarks of ResMed Ltd and registered in U.S. Patent and Trademark Office.
©2007 ResMed. 1012117/2 07 12
Global leaders in sleep and respiratory medicine www.resmed.com
PRODUCT CODES
VPAP Auto 26301
HumidAire 3i 33906
TECHNICAL SPECIFICATIONS
Performance
Operating pressure range: 3 to 20 cm H
2
O
Maximum single fault pressure: 30 cm H
2
O
Dynamic pressure characteristics
VAuto mode: 4 to 20 cm H
2
O (Pressure support: 0 to 10 cm H
2
O)
S mode: IPAP: 5 to 20 cm H
2
O; EPAP: 3 to 20 cm H
2
O
CPAP mode: 4 to 20 cm H
2
O
Sound pressure level
< 24 dB (tested in accordance with the requirements of ISO 17510-1:2002)
Data storage
Data includes usage, leak, pressure, AHI, AI, tidal volume, respiratory rate
and minute ventilation
Via LCD menu: 365 sessions of summary data
Via ResLink: 365 sessions of summary data, 30 sessions of detailed data
Via ResScan Data Card: 180 sessions of summary data, two sessions of
detailed data
Dimensions (L x W x H): 4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power supply
Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)
Environmental conditions
Operating temperature: 41°F to 97°F (+5°C to +36°C)
Operating humidity: 10–95% non-condensing
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) for residential, commercial and light industry
environments according to IEC 60601-1-2.
Air filter
Two-layered, powder-bonded, polyester non-woven fiber
IEC 60601-1 classification
Class II (double insulated), Type CF
Note: The manufacturer reserves the right to change these specifications
without notice.
VPAP AUTO FEATURES SUMMARY
Modes of operation
VAuto
S (Spontaneous)
•CPAP
Product features
Simplified standard setting menu
Vsync
automatic leak management
TiControl
(Ti max/min)
Adjustable breath trigger/cycle
Adjustable exhalation comfort
12V DC input (via converter)
Auto-adjusting bilevel pressures
Compact S8 design
AutoSet technology
Smooth pressure delivery
Heated humidification
Optional, fully integrated
Data storage
Usage and Treatment Summary (day, week, month, six-month, year)
Usage profile
Summary data storage
Detailed data storage
Remote settings adjustment
Data transfer
Direct download to PC
ResLink via SmartMedia card
ResScan Data Card