ResMed 60561 Sleep Apnea Machine User Manual


 
7
English
Troubleshooting
Possible reason Solution
Pillows won’t seal properly, are uncomfortable or cause red marks
Pillows may have
been fi tted incorrectly,
incorrectly adjusted, or
the headgear is too tight.
Carefully follow instructions in “Fitting your Swift LT”.
Check headgear is not over-tightened.
Check ResMed logo on top of pillows is facing
outwards.
Inner wall position of one/
both pillows is incorrect.
Squeeze pillow sides to release inner wall. The
correct position is shown in “Reassembling your
Swift LT”.
Pillows are wrong size. Talk to your clinician.
Pillows may be dirty. Clean pillows according to instructions in “Cleaning
your Swift LT in the Home”.
Nasal pillows system is too noisy
System is assembled
incorrectly.
Reassemble the Swift LT. Make sure the pillows are
fully inserted into the frame.
Before using your Swift LT
WARNINGS
The vent holes must be kept clear.
The mask should not be used unless the CPAP system is turned on and
operating properly.
Follow all precautions when using supplemental oxygen.
Oxygen fl ow must be turned off when the fl ow generator is not operating,
so that unused oxygen does not accumulate within the fl ow generator
enclosure and create a risk of fi re.
At a fi xed fl ow rate of supplemental oxygen fl ow, the inhaled oxygen
concentration varies, depending on the pressure settings, patient breathing
pattern, mask, point of application and leak rate.
The technical specifi cations of the mask are provided for your clinician to
check that they are compatible with the fl ow generator. If used outside
specifi cation or if used with incompatible devices, the seal and comfort of
the mask may not be effective, optimum therapy may not be achieved, and
leak, or variation in the rate of leak, may affect the fl ow generator function.
Discontinue using the Swift LT if you have ANY adverse reaction to the use
of the mask, and consult your physician or sleep therapist.
As with all masks, some rebreathing may occur at low CPAP pressures.
Refer to your fl ow generator manual for details on settings and operational
information.
Remove all packaging before using the mask.