Mask and headgear are not positioned
correctly.
Adjust position of mask and headgear.
Incorrect air tubing selected. If you are using the SlimLine or Standard air tubing ensure
that you have the correct air tubing selected via the menu.
Pressure rises inappropriately
Talking, coughing or breathing in an
unusual manner.
Avoid talking with a nasal mask on, and breathe as normally
as possible.
Mask cushion is buzzing against the
skin.
Adjust the headgear.
Displays error message: High temperature fault, refer to user manual
Device has been left in a hot
environment.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Air filter is blocked. Replace your air filter. Disconnect the power cord and then
reconnect it to restart the device.
Air tubing is blocked. Check your air tubing and remove any blockages.
Disconnect the power cord and then reconnect it to restart
the device.
Humidifier setting is too high, resulting
in accumulation of water in the air
tubing.
Turn the humidifier setting down and empty the water from
the air tubing.
Displays error message: Check ResMed 30/90W Power Supply Unit and fully insert the
connector
The DC plug is partially inserted into the
back of the device or inserted too
slowly.
Fully insert the DC plug.
A non-ResMed power supply unit is
connected to the device.
Remove the power supply unit and replace with a ResMed
power supply unit.
The power supply unit is being covered
by bedding.
Make sure that the power supply unit is free from bedding,
clothes or other objects that could cover it.
Displays error message: Tube blocked, please check your tube
Air tubing is blocked. Check your air tubing and remove any blockages.