ReliOn HEM-780REL Blood Pressure Monitor User Manual


 
27
SPECIFICATIONS
Model: ReliOn
®
HEM-780REL
Display: LCD Digital Display
Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40-180/min.
Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Inflation: Automatic by electric pump
Deflation: Automatic pressure release valve
Rapid Air Release: Automatic exhaust valve
Pressure Detection: Capacitive pressure sensor
Measurement Method: Oscillometric method
Pulse Wave Detection: Capacitance type pressure sensor
Power Source: AC Adapter (or 4 “AA” batteries)
Battery Life: Approx. 1500 uses when used once a day
for 2 minutes with 4 new alkaline batteries
Operating Temperature
/Humidity: 50ºF to 104ºF (10ºC to 40ºC) /30 to 90% RH
Storage Temperature
/Humidity: -4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
Main Unit Weight: Approximately 12 1/2 oz. (355 g)
not including batteries
Main Unit Dimensions: Approx. 6 3/32" (l) x 5 5/32" (w) x 3 5/16" (h)
(155 mm x 131 mm x 84 mm)
Cuff Size: Approximately 6" (w)(152 mm)
Cuff tube 23 5/8" (l) (600 mm)
Cuff Circumference: Fits arm circumferences 9" to 17"
(228mm to 432mm)
Contents: Main Unit, Easy Wrap Cuff, AC Adapter,
Instruction Manual, 4 “AA” Batteries,
and Storage Case
UPC Code: 6 8113189539 2
NOTE: These specifications are subject to change without notice.
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 26