M3860A/M3861A HEARTSTART FR2+
6-12
Philips Medical Systems
The highlighted top line is labeled MANUAL CHARGE, with an arrowhead
pointing to the upper Option button.
If the ECG display shows that, in your expert clinical judgment, the patient
has a shockable rhythm, press the upper Option button (
MANUAL CHARGE).
The HeartStart FR2+ will immediately charge for shock delivery.
As soon as charging begins, the screen message changes to
CHARGING,
STAND CLEAR, and the label for the arrowhead pointing to the upper Option
button changes to
MANUAL DISARM.
The FR2+ beeps while it is charging. When the beeping changes to a
continuous tone and the Shock button light flashes, press the Shock button
to deliver a shock. However, if the ECG display shows that the patient’s
rhythm has changed to a non-shockable rhythm, press the upper Option
button to disarm the HeartStart FR2+.
After shock delivery, the HeartStart FR2+ returns to the initial advanced
mode screen. To return to non-manual, AED mode operation, turn the FR2+
off by pressing the On/Off button. Then turn the FR2+ on by pressing the
On/Off button again.