Philips CM-200 Microscope & Magnifier User Manual


 
9.1 Press STIG button to open the STIGMATOR CONTROL page.
9.2 Select Condense Aperture.
9.3 Use the MULTIFUNCTION X/Y knobs, one at a time, such that the beam spot circle has a fixed
center when the beam intensity is changed.
10. High Resolution Images
10.1 Bring in the small screen into viewing chamber.
10.2 TV on.
10.3 Amplifier on.
11. Load Films
11.1 Press CAMERA AIR.
11.2 Put film in.
11.3 PUMP till to IGP less than 10.
11.4 Note films, along with film box should be put into a vacuum for more than 24 hours to degas.
Note: reset the negative numbers
6