31
English
32
English
Q & A (continued)
Q
Since
the
mark was displayed, I retried the measurement
by setting my arm on the unit properly. But the indicated value
is still higher than usual.
A
The
mark is just a guide for the proper measurement method.
(See page 20.)
If you correct your posture when the mark is displayed but the
measured value is still higher or lower than usual, retry the
measurement.
The following causes are suspected.
Did you measure properly (in proper posture)?
Measure as indicated on pages 15–17.
Measured immediately after smoking.
Measured when irritated or emotional.
Measured in cold environments
For persons who have a physical constitution with less fluctuation
in the pulse rate.
If you are concerned about the results, retry the measurement after
staying quiet for 4 to 5 minutes.
•
•
•
•
•
•
Q When
should I measure my blood pressure?
A We recommend three measurements a day.
First ........ Upon
Waking (Because, with almost no influence from
body movement, it’s the blood pressure is closest to that
at night.)
Second ... Before Lunch (Because it provides blood pressure values
close to the highest of the day.)
Third ....... At Bedtime (So as to obtain data close to your average
blood pressure values during the day.)
This way, by taking measurements morning, noon and night, you can
use the unit to learn the patterns of and monitor your blood pressure.
Use of the AM and PM average systolic blood pressure comparison
function can aid in the detection of a blood pressure pattern known
as “early morning surge” in which blood pressure rises rapidly in
the morning. (See page 27.)
•
Care and Maintenance
Do not attempt to disassemble, repair or modify the unit.
Doing so may cause fire or cause the unit to malfunction. It may also
lead to injury.
Do not apply excessive force to the unit or drop the unit.
Doing so may cause damage.
Do not insert dust or foreign objects into the unit.
Doing so may result in damage.
If the unit is stored at temperatures below the freezing point, do not use it
immediately.
Leave it in a warm place for at least an hour before use.
If the unit is not allowed to warm up, it may not pressurize.
Do not touch the unit when measurement is in progress.
Doing so may cause incorrect readings.
If the unit becomes dirty, clean it with a soft cloth moistened with warm
water or soapy water.
(Do not use alcohol, benzine, or paint thinner.)
Use of such chemicals may result in cracking or discoloration.
When the unit will not be used for a long time (30 days or more), be sure
to remove the batteries.
Otherwise, the batteries may leak and damage the unit.
The pressure cuff cannot be washed.
For storage, avoid high temperatures, high humidity, and direct sunlight.
Shortage under such conditions may result in damage.
•
•
•
•
•
•
•
•
31-32 2008/01/15 14:54:31