Omron M7 Blood Pressure Monitor User Manual


 
12
M7 English
6 Storage
1 Slightly fold air tube and insert in cuff.
Note: do not disconnect air tube.
2 Put cuff and main unit in storage case.
Remove the batteries if you do not intend to use the OMRON M7 for 3 months or more.
Store the OMRON M7 in the carton in a dry, closed place with a temperature between -20°C and 60°C.
M
DATE/TIME
SET
START
SYS
mmHg
DIA
mmHg
PULSE
/min.
M7
Indication Error Cause Rectification
Incorrect inflation Movement Repeat measurement.
If necessary, hold start
button to inflate manually
Correct Movement, sound Follow instructions, repeat
measurement Incorrect fastening cuff measurement
not possible Operation start button
for more than 3 seconds
Cuff leaks air Fit new cuff
Cuff pressure too high Follow instructions, repeat
measurement
7 Error: causes and rectifications