11
EN
4.
Troubleshooting and Maintenance
4.1
The Icons and Error Messages
Error Display Cause Remedy
Device stabilization in process.
Wait until stops flashing.
Battery is low.
Replace the battery.
(Refer to section 4.3)
Probe cover is not attached correctly.
Attach the probe cover again until stops flashing.
Measurement before device stabilization.
Wait until stops flashing.
The device is showing a rapid ambient
temperature change.
Allow the thermometer to rest in a room for at least
30 minutes at room temperature: 10°C and 40°C
(50°F - 104°F).
The ambient temperature is not within
the rangebetween 10°C and 40°C
(50°F - 104°F).
Allow the thermometer to rest in a room for at least
30 minutes at room temperature: 10°C and 40°C
(50°F - 104°F).
Error 5-9, the system is not functioning
properly.
Remove the battery, wait for 1 minute and repower it. If
the message reappears, please contact the OMRON
retail outlet or distributor for having the device checked.