Omega Engineering HH505 Thermometer User Manual


 
1
HH 504 / HH 505 CALIBRATION PROCEDURE
Note: The following calibration procedure should perform only by qualified technicians
who have access to the items as following.
Equipment: The class of calibrator had better 10 times greater than the measured meter.
1. Turn off the meter then set the adjusted jumper to J1.
2. Insert the 383.00Ω calibrator to the T1 then turn on the meter, the main
display reading reads 1830.0 then press " ENTER " key.
3. The second display is stabilized (about 5sec.), then press " ENTER " key
when the meter auto power off then set the jumper back to J3 position.
4. Turn on the meter then Insert the 100.00Ω calibrator. The display reading
reads 0°C if the calibration procedure is right.
Pt385 Pt3926
-200 18.521Ω 16.996Ω
-100 60.256Ω 59.479Ω
0 100.000Ω 100.000Ω
100 138.505Ω 139.272Ω
200 175.856Ω 177.362Ω
300 212.052Ω 214.275Ω
400 247.092Ω 250.018Ω
500 280.977Ω 284.591Ω
600 313.708Ω 317.994Ω
700 345.280Ω
800 375.700Ω
383.00W CALIBRATOR
+
RW1
RW1=RW2=RW3
RW=WIRERESISTANCE
G
+
_
_
RW2
383.00W
G
RW3