Nikon 55i Microscope & Magnifier User Manual


 
En
This brochure is printed on recycled paper made from 40% used material.
WARNING
TO ENSURE CORRECT USAGE, READ THE CORRESPONDING MANUALS
CAREFULLY BEFORE USING YOUR EQUIPMENT.
Specifications and equipment are subject to change without any notice or obligation
on the part of the manufacturer. December 2003. ©2003 NIKON CORPORATION
NIKON CORPORATION
Yokohama Plant
NIKON CORPORATION
Instruments Company
ISO 9001
Accredited by the
Dutch Council for
Accreditation
ISO 14001
Printed in Japan (0312-13)T
Code No. 2CE-MSHH-1
www.nikon.com/
Fuji Bldg., 2-3, Marunouchi 3-chome, Chiyoda-ku,
Tokyo 100-8331, Japan
NIKON CORPORATION
NIKON INSTECH CO., LTD.
Parale Mitsui Bldg.,8, Higashida-cho, Kawasaki-ku,
Kawasaki, Kanagawa 210-0005, Japan
phone: +81-44-223-2167 fax: +81-44-223-2182
www.nikon-instruments.jp/eng/
www.i-series.com
NIKON INSTRUMENTS (SHANGHAI) CO., LTD.
CHINA phone: +86-21-5058-5055 fax: +86-21-5058-5060
NIKON SINGAPORE PTE LTD
SINGAPORE phone: +65-6559-3618 fax: +65-6559-3668
NIKON MALAYSIA SDN. BHD.
MALAYSIA phone: +60-3-78763887 fax: +60-3-78763387
NIKON INSTRUMENTS EUROPE B.V.
P.O. Box 222, 1170 AE Badhoevedorp, The Netherlands
phone: +31-20-44-96-222 fax: +31-20-44-96-298
http://www.nikon-instruments.com/
NIKON FRANCE S.A.
FRANCE phone: +33-1-45-16-45-16 fax: +33-1-45-16-00-33
NIKON GMBH
GERMANY phone: +49-211-9414-0 fax: +49-211-9414-322
NIKON INSTRUMENTS S.p.A.
ITALY phone: + 39-55-3009601 fax: + 39-55-300993
NIKON AG
SWITZERLAND phone: +41-43-277-2860 fax: +41-43-277-2861
NIKON UK LTD.
UNITED KINGDOM phone: +44-20-8541-4440 fax: +44-20-8541-4584
NIKON INSTRUMENTS INC.
1300 Walt Whitman Road, Melville, N.Y. 11747-3064, U.S.A.
phone: +1-631-547-8500; +1-800-52-NIKON (within the U.S.A.only) fax: +1-631-547-0306
http://www.nikonusa.com/
NIKON CANADA INC.
CANADA phone: +1-905-625-9910 fax: +1-905-625-0103
* Monitor images are simulated.
Company names and product names appearing in this brochure are their registered trademarks or trademarks.
Clinical & Laboratory Microscopes ECLIPSE 50i/55i
. . . magine microscopy in perfect comfort
/
/
200
330
60
404
208
60
I
O
FIL
486
330
200
208
Dimensional Diagrams
Configured with binocular tube B Configured with ergonomic tube
Unit: mm
Specifications
Main body 50i 55i
Magnification 10–1500X
Optical system CFI60 Infinity Optical System
Coarse/fine focusing Fine: 0.1mm per rotation, Coarse: 13.8mm per rotation, Minimum reading: 1 micron
Coarse motion torque adjustable, Refocusing function (with stopper)
Illumination 6V-30W halogen lamp White LED array
100-240V (worldwide voltage)
AC adapter (100-240V) or optional lithium battery (7.4V)
Built-in filter ND8 LA60 color balance filter
Eyepiece tube Binocular tube B (for F.O.V. 22mm)
Trinocular tube "F" UW (for F.O.V. 22mm/25mm, observation/photo: 100/0, 0/100)
Trinocular tube "T" UW (for F.O.V. 22mm/25mm, observation/photo: 100/0, 20/80, 0/100)
Ergonomic binocular tube (for F.O.V. 22mm, inclination:10-30°, extension: 40mm); DSC port: 100/0, 50/50 (optional)
Eyepiece lens 10X (F.O.V.
:
22mm), 10X M photo mask (F.O.V.: 25mm), 12.5X (F.O.V.
:
16mm), 15X (F.O.V.
:
14.5mm),
UW 10X (F.O.V.
:
25mm), UW 10X M photo mask (F.O.V.
:
25mm)
Nosepiece Sextuple nosepiece
Ergo-View cytodiagnostic unit (Motorized changeover 10X to 40X with hand switch, Stamp type marking
Automatic illumination adjustment with Ergo-View
Stage Super-hard Alumite coated surface, Stay-in-position stage handle, Stage handle height and tension adjustable
Rectangular 159mm X 243mm surface stage, 78mm X 54mm cross travel (x-y movement)
1-slide or 2-slide specimen holder available (option)
Condenser focusing stroke 27mm
Intermediate accessories Epi-fluorescence illuminator (4 filter positions), Magnification module, Eye-level riser, Double port
Teaching head
Observation method Brightfield, Epi-fluorescence
Darkfield, Phase contrast, Simple polarizing
ECLIPSE i-Series Website
www.nikon-i.com
Please contact Nikon for a handy pamphlet listing compatible
accessories, including objectives and epi-fluorescence filters.
Nikon has reduced the amount of chromium, cadmium and lead used in the Eclipse-i Series
to an absolute minimum to diminish its environmental impact.
Microscopy images courtesy of :
1Nihon University Itabashi Hospital
2Naoyuki Miyokawa, M.D., Ph.D., Associate
Professor, Dept. of Surgical Pathology,
Asahikawa Medical College Hospital.