15..
Replacing the Air Filter
The air filter prevents extraneous substances from entering the mist generated by the
device. Replace the air filter after about 20 hours of use, or when the filter turns gray.
There is a package of 5 replacement air filters supplied with the nebulizer. Additional
filters can be purchased from Graham-Field.
1. Remove the top cover from the device.
2. Lift the air filter holder from the top cover.
3. Remove the filter from the air filter holder.
4. Insert a new filter into the air filter holder.
5. Replace the air filter holder on the top cover.
Caution: Do not use cotton or other material as a filter.
Caution: Do not wash or clean the filter.
Caution: Do not reuse a dirty filter. Do not operate the nebulizer without a filter; nebulizer damage can result.
STORAGE
Keep the device and accessories dry. Avoid exposure to direct sunlight, high moisture, and dust.
Caution: If the device is to be stored long-term, remove the battery.
The battery's storage environment must be between -4°F and 95°F (-20°C and 35°C).
The device's storage environment must be between -13°F and 140°F (-25°C and 60°C) and 10% - 95% RH.
6700-INS-LAB-RevA08.indd 15 1/27/09 4:58 PM