LifeSource UA-774 Blood Pressure Monitor User Manual


 
LifeSource Health Line (Toll-Free): 1-888-726-9966
E-17
Model . . . . . . . . . . . . . . . . . . .UA-774
Type . . . . . . . . . . . . . . . . . . . . .Oscillometric
Display . . . . . . . . . . . . . . . . . .Digital character height
. . . . . . . . . . . . . . . . . . . . . . .Pressure (16.0 mm)/
pulse (10.0 mm)
. . . . . . . . . . . . . . . . . . . . . . .displayed simultaneously
Memory . . . . . . . . . . . . . . . . . . . .30 readings per person for two
persons
Measurement range . . . . . . . . .Pressure: 20 mmHg to 280 mmHg
Pulse: 40 pulses to 200 pulses
per minute
Accuracy . . . . . . . . . . . . . . . . .Pressure: +3 mmHg or +2%,
whichever is greater
Pulse: +
5%
Pressurization . . . . . . . . . . . .Automatic, using micropump
Depressurization . . . . . . . . . .Automatic constant speed
Power source . . . . . . . . . . . . .4 type “AA” (1.5 volt) alkaline
batteries (not included) or
AC Adapter (A&D Model TB:181)
Battery life . . . . . . . . . . . . . . .Approximately 6 months with
one daily measurement
Operating environment . . . .50ºF to 104ºF (10ºC to 40ºC)
Less than 85% relative humidity
Storage environment . . . . . .14ºF to 140ºF (-10ºC to 60ºC)
Less than 95% relative humidity
Dimensions . . . . . . . . . . . . . . .Length: 4.3” (110 mm)
Width: 5.8” (147 mm)
Height: 2.5” (64 mm)
Weight . . . . . . . . . . . . . . . . . . .10.6 oz. (300 g) without batteries
Blood pressure measurements determined by the UA-774 are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method
within the limits prescribed by the American National Standards Institute
(ANSI/AAMI SP-10) for electronic or automated sphygmomanometers.
SPECIFICATIONS
TROUBLESHOOTING
Problem
Nothing appears in
the display, even
when the power is
turned on.
The cuff does not
inflate.
The unit does not
measure.
Readings are too
high or too low.
Other
Probable cause
Batteries are drained.
Battery terminals are
not in the correct
position.
Battery voltage is too
low. LOW BATTERY
MARK ( ) blinks.
[If the batteries are drained
completely, the mark does
not appear.]
The cuff is not fastened
properly.
You moved your arm or
body during the
measurement.
The cuff position is not
correct.
You are using the wrong
size cuff.
The value is different
from that measured at a
clinic or doctor’s office.
NOTE: If the actions described above do not solve the problem, contact
A&D Medical or Auto Control Medical at the numbers listed at the
bottom of these pages. Do not attempt to repair the device yourself.
Corrective action
Replace all batteries with new
ones.
Reinstall the batteries with
negative and positive terminals
matching those indicated in the
battery compartment.
Replace all batteries with new
ones.
Fasten the cuff correctly.
Make sure you remain very still
and quiet during the
measurement.
Sit comfortably and still. Make
sure the cuff is at the same level
as your heart.
See Pg. E-5 “Selecting the
Correct Cuff.”
See Pg. E-14 “Why Measure
Blood Pressure at Home.”
Remove the batteries. Place
them back properly and try the
measurement again.
E-16
In Canada (Toll-Free): 1-800-463-5414