Leica M80 Microscope & Magnifier User Manual


 
Leica M-series User Manual 7
Safety Regulations (cont'd.)
Light sources: safety regulations
Protective measures of the manufacturer:
UV protection screen in front of the speci-
men plane prevents the user from looking
directly into the UV rays.
Dummy filter carriers in the free positions
of the rapid filter changer prevent direct UV
radiation from reaching the eyes.
UV filters are installed in the observation
beam paths to protect the eyes.
The stray-light protection on the lamp
housing prevents irradiation of the hands.
Warning
UV radiation could damage the eyes.
Therefore:
Never look into the light spot on the speci-
men plane without a UV protection screen.
Never look into the eyepieces if no excita-
tion filter is in the beam path.
Fill empty filter positions with dummy filter
carriers (e.g. M165 FC).
Do not select a white, strongly reflective
background for the specimen.
Supply unit
Always unplug the supply unit from the power
supply:
When installing and disassembling the
lamp housing
Before opening the lamp housing
When replacing the high-pressure mercury
lamp and other parts, such as the heat-
absorbing filter or the collector
During maintenance work on the supply
unit