Leica M125 B Microscope & Magnifier User Manual


 
Leica M series User Manual 10
Safety Instructions (continued)
Light sources: safety regulations
Light sources pose a potential risk from
glare. Therefore, lamps have to be operated
in closed housings and in installed condi-
tion.
Never look directly into the beam path
(blinding hazard).
Do not select a white, strongly reflective
background for the specimen.
External power supply for the focusing
column
(Leica M165 C B)
Permitted power supply:
SINPRO SPU46-110
Specifications:
Input: 100-240 V AC
47-63 Hz
1.2 A
Output: 33 V DC
1.51 A
For indoor use only.
External power supply for TL5000 Ergo
Permitted power supply:
SINPRO SPU130-110
Specifications:
Input: 100-240 V AC
47-63 Hz
3.2 A
Output: 33 V DC
3.93 A
max. 130 W
For indoor use only.
•
Use only the power supply specified
above. Other power supplies must not
be used. If the original power supply fails or
is damaged, it must be replaced. Repair is not
permitted. Original power supplies are avail-
able from your Leica branch office or Leica
dealer.