34
Operating Manual V1.1 – 05/2007
5. Working with the Instrument
5.4.4 Installing and Removing the Blade Holder
Fig. 19
• Before the blade holder can be replaced, it
must be brought into a 45° inclined position.
To do so, insert the size 3 Allen key sideways
into the blade holder through the hole (8) and
rotate it 45° clockwise.
Turning the screw (9) counterclockwise
releases the blade holder, allowing it to be
replaced.
• The installation is carried out in reverse order.
Cleaning the blade holder
To clean the blade holder after removal, spray it
with alcohol. It can then be wiped down with a
piece of cellulose and placed on a cellulose tow-
el to dry completely.
8
For quality and service reasons, the
blade holder (7) is available as a com-
plete unit only.
9
Always remove the blade BEFORE in-
stalling or removing the blade holder!
7
7