Invacare XPO100B Oxygen Equipment User Manual


 
SECTION 8—TROUBLESHOOTING GUIDE
Invacare XPO
2
42 Part No 1148112
BATTERY TOO HOT/COLD ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.
AND
Flow indicators 1
& 4 are
illuminated.
Internal Battery has become
too hot, or too cold, during
operation.
1. Unplug chargers if
connected.
2. Allow unit to cool
down to less than 95ºF,
or warm up to 50ºF,
before resuming
charging or turning Off
the unit and turning the
unit On again.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider.
STUCK BUTTON ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.
AND
Flow indicators 1
& 5 are
illuminated.
A stuck button is being
detected upon power up.
1. Turn Off the unit
2. With the unit turned
Off, press each control
button looking for a
stuck button
3. Turn On the unit
while ensuring that you
are not continuing to
hold the Power On/Off
button once the lights
begin to illuminate.
4. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
ALARM/WARNING:
SYMPTOM PROBABLE CAUSE SOLUTION