Invacare R102 Mobility Aid User Manual


 
OPERATING THE PATIENT LIFT
Electric Portable Patient Lift 24 Part No. 1145810
Secondary Emergency Release
NOTE:Forthisprocedure,refertoFIGURE 3.3.
NOTE:Usetheprimaryemergencyrelease
first.Thesecondaryemergencyreleaseisonlya
backuptotheprimaryemergencyrelease.
Incaseswheretheprimaryreleaseiseither
notfunctioningorunreachable,a
secondaryemergencyreleasemaybeused.
Toactivatethesecondaryrelease,
pullup
ontheEMERGENCYringandpushdown
ontheboomatthesametime.
FIGURE 3.3 Secondary Emergency Release
Performing an Emergency Stop
NOTE:Forthisprocedure,refertoFIGURE 3.4.
PresstheRED/ORANGEemergencybuttononthecontrolboxtostoptheboomassembly
andpatientfromraisingorlowering.
RotatetheRED/ORANGEemergencystopbuttonclockwisetodisengagetheemergency
stop.
FIGURE 3.4 Performing an Emergency Stop
Charging the Battery
NOTE:Forthisprocedure,refertoFIGURE 3.5onpage25.
NOTE:Invacarerecommendsthebatteryberechargeddailytoprolongbatterylife.
NOTE:Anaudiblealarmwillsound(hornwillbeep)whenbatteryislow.
1. Liftuponthehandleonthebackofthebattery.
2. Liftthebatteryupandout
ofthecontrolbox.
Pull UP on
EMERGENCY
Ring
Push DOWN
on Boom
Control Box
Press IN to
Stop Boom
Rotate CLOCKWISE to
Disengage Emergency Stop