Invacare IRC5LXO2, IRC5LX Respiratory Product User Manual


 
8
SPECIFICATIONS
Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pressure:
120 VAC +10% (132 VAC/108 VAC),
60 Hz
4.0 A
50 dB Average
Up to 6,000 ft. (1828 meters) above
sea level without degradation of
concentration levels. From 6,000 ft.
(1828 meters) to 13,129 ft. (4000
meters) below 90% efficiency.
ALL 5LXO2/5LX MODELS
95.6% to 87% at 1/2 to 5 L/min.
* Stated concentration levels
achieved after initial warm-up
period (approximately 30 minutes).
5
+ 0.5 psi.
(34.5 + 3.45 kPa.)
SPECIFICATIONS
SensO
2
® Alarm Thresholds:
O2 Purity
Below 73%
73% to 85%
Over 85%
Indicators
RED Indicator Light -
Continuous Audible Alarm
Sieve-GARDCompressor
Shutdown
YELLOW Indicator Light
GREEN Indicator Light
O
2
Label
Symbol