Invacare JLS5 Scale User Manual


 
Jasmine™ Scale JLS5 6 Part No 1152408
Replacing the Battery
NOTE:Forthisprocedure,refertoFIGURE 4.
NOTE:Thescaleispoweredbya9voltalkalinebatterythatshouldprovideapproximately1500readingsbefore
needingreplacement.
Whenbatteryreplacementisneeded,LOBATwillappearonthedisplay.Performthefollowing:
1. Slidethebatterydooropeninthedirectionofthearrow.
2. Removeexistingbattery.
3. InstalltheNEWbattery.
4. Reinstallthebatterydoor.
FIGURE 4 Replacing the Battery
Calibrate the Jasmine Scale
NOTE:Forthispr ocedur e,refertoFIGUR E 5onpage7.
NOTE:TheJasmineScalewill bepre calibratedatthefactory withtheloadcell.Shoulditbenecessarytorecalibrate
thescale,followtheinstructionsoutlinedbelow.
1. ThepatientandtheslingMUSTberemov edfromthescaletoproperlycalibratetheJas mineScale.Refer
toOperationintheJasminePatientLiftOwner’sManual,InvacareP/N1150704,forfurtherinstructions.
2. Forslingdetachmentinstructions, refertoLiftingthePatientintheJasminePatientLiftOwner’sManual,
Inv acareP/N1150704.
3. Turnthescaledisplaypow eronbypressingtheON/OFFkey.
4. Rotatethe SwivelBaruntilthebackofth escaledisplayisvisible.
5. Insertthetipofapaperclipor similartoolintoth esmallholeonthebackof
thescaledispla yandpress
theCALbutton.
NOTE:OncetheCALbuttonispr essed,“CAL1”willbeseeninthedisplaywindow.
6. PresstheUNITSkey.ThedisplayreadsCAL2.
7. PresstheLOCK/UNLOCKkey.ThedisplayreadsUNLD.
8. PresstheZEROkey.Thedisplayreads LOAD.
9. Hang200poundsofcalibratedweightfromthescaleandpresstheZEROkey.
10. PresstheZEROkey.Thedisplayreads DONE.
Battery
Door
9V Battery
Assembled
Scale
Arrow