Invacare IRC5P Respiratory Product User Manual


 
SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1143482 13 Perfecto
2
Series
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input: Perfecto
2
5: 3 A
Sound Level: Perfecto
2
5: 44 ± 2 dBA Average
Altitude: Perfecto
2
5 - Up to 6,000 ft. (1828 meters)
above sea level without degradation of con-
centration levels. From 6,000 ft. (1828
meters) to 13,129 ft. (4000 meters) below
90% efficiency.
*Oxygen Output
Concentration Levels:
All IRC5PO2/IRC5P models 95.6% to 87%
at 0.5 to 5 L/min
Maximum Outlet Pressure: Perfecto
2
5: 5 psi ± 0.5 psi. (34.5 kPa ± 3.45
kPa.)
Flow Range: Perfecto
2
5 - 0.5 to 5 L/min
For flowrates less than 1 L/min, we recom-
mend the use of the Invacare Pediatric Flow-
meter IRCPF16 Accessory.
Low Flow Alarm: Perfecto
2
5 - 0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected).
Average Power
Consumption:
Perfecto
2
5: 280 W (Operating @ 3 L/min)
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:
0.7 L/min
Filters: Cabinet, Outlet HEPA and Compressor
Inlet