Invacare IRC5LX Oxygen Equipment User Manual


 
SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1118353 15 Platinum
Series
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity
Exhaust: Platinum 5 and XL: Less than Ambient
+ 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F
(+25°C)
Oxygen Output Temperature:
Platinum 5 and XL: Less than Ambient + 4°F
(+2°C)
Platinum 10: Less than Ambient + 8°F
(+4.4°C)
Cabinet: Impact Resistant flame-retardant plastic cab-
inet that conforms to UL 94-V0
Regulatory Listing: Double Insulated Product ETL certified
complying with UL 1097 ETL certified to
CSA C22.2 No. 68
Electrical: No extension cords.
Placement: No closer than three inches from any wall,
furniture, draperies, or similar surfaces.
Tubing: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft
high flow tubing with high flow cannula at all
flow rates.
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity