Invacare HF2POSTD Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
18 Part No 1148114
ƽ WARNING
The fillport cover on the connector fillport and the post
valve fillport MUST be replaced after filling and whenever
not in use. If either of the fillport covers are missing
contact your healthcare provider before using.
17. Place the fillport covers onto the connector fillport (located
on the compressor) and post valve fillport (located on the
cylinder) (FIGURE 4.7).
18. Install a CGA870 post valve compatible conserver or
regulator according to the manufacturer’s instructions.
19. Open the toggle valve by turning counterclockwise and
examine the conserver or regulator gauge to ensure
cylinder is full (needle pointing into GREEN area). If the
cylinder is not full, perform the following:
Repeat Filling the Cylinder on page 13.
If cylinder is not full after second attempt to fill, contact
the provider.
FIGURE 4.7 Connector Fillport
Connector
Fillport Cover
Fillport Cover
Toggle
Valve