Invacare HF2POST9 Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
12 Part No 1148114
FIGURE 4.1 External Examination
Checking Cylinder Pressure
ƽ WARNING
NEVER use tools of any kind to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise, injury
or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 4.2 on page 14.
1. Install a conserver or regulator onto the cylinder according
to the device manufacturer instructions.
DO NOT REMOVETHIS PRODUCT LABEL
Not for Commercial Resale
Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford, FL 32773
1-800-832-4707
Part Number 1125090 Rev B - 11/04
OXYGEN
2
This cylinder is to be used with the Invacare Venture
HomeFill system.This cylinder contains
93 + 3%
oxygen produced by an oxygen concentrator using
the pressure swing adsorption process.
DO NOT handle cylinder or use contents until you are
professionally trained, including emergency procedures.
WARNING
For medical applications. Use only
as directed by a
licensed
practitioner. Uninterrupted use of
high concentrations
of oxygen
over a long duration, without
monitoring its effect on
oxygen
content of arterial blood, may be harmful.
Oxygen Cylinders filled by the Venture HomeFill
shall be used for personal use only. "Not to be
filled for resale or use by professional users".
!
!
DANGER
Improper filling or abuse of this cylinder, or failure to follow
these safety instructions may cause serious bodily injury or
death. High pressure oxidizing gas vigorously accelerates
combustion.
DO NOT alter or modify cylinder or related components.
DO NOT use Oil or Grease. Keep cylinder, cylinder valve, all
equipment and connections free of oil and grease to avoid
violent ignition resulting in serious bodily injury or death.
Use in accordance
with Venture HomeFill Operator’s
Manual.
Open valve slowly.
Close valve after each use
and when empty.
S
ecure cylinder during s
torage and use.No smoking in cylinder
area.
Keep away from heat, flame and spark. Keep out of
reach of children. DO NOT drop.
Transfilling of this gas is performed by Venture HomeFill Only.
Cylinders that have been refinished and/or exposed to
elevated temperatures must be hydrostatically tested before
filling.
Cylinder temperature
should not exceed 125°F (52°C).
Aluminum cylinders subject
to the action of
fire or heated to
temperatures in excess of 350°F
must be
withdrawn from
service and condemned by trained
personnel. DO NOT use
caustic paint strippers.Valve and safety relief devices must be
removed and replaced by trained and authorized
personnel. DO NOT alter or change valve assembly.
+