SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1143247 13 Oxygen Conserving Cylinder
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
ƽ WARNING
NEVER use tools of any kind to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise, injury
or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 4.2 on page 14 and
FIGURE 4.3 on page 15.
1. Examine the cylinder gauge. If the cylinder pressure is less
than 1500 psig., proceed to STEP 2 (FIGURE 4.3 on
page 15).
NOTE: If the cylinder pressure is greater than 1500 psig., DO
NOT attempt to top off this cylinder. It may not fill.
2. Set the regulator flow dial on the cylinder to OFF (FIGURE
4.3 on page 15).
3. Remove the cylinder and connector fillport covers (if
present).
CAUTION
DO NOT connect the outlet (cannula) connection to the
HomeFill unit, otherwise the conserving cylinder will not
work properly.