5.7 Troubleshooting
FAULT POSSIBLE CAUSE REMEDY
No motor sound audible/the lift
does not move
Battery is empty or not sufficiently charged
Battery contacts are dirty
Battery is not inserted correctly
Cable for hand control unit is damaged.
Recharge battery fully
Clean battery contacts
Insert battery correctly
Contact a qualified technician.
Lift can only be raised Battery not sufficiently charged Recharge battery fully
Lift stops when rising Maximum load exceeded Observe maximum load
LED on power socket does not
light up during charging
Battery defective
Charger not inserted correctly
Charger defective
Battery is already fully charged
Replace battery
Insert charger correctly
Replace charger
-----
Device slips Suction cups damaged Replace suction cups
LED on hand control unit lights up
during operation
Battery not sufficiently charged
Electronics defective
Recharge battery fully
Contact a qualified technician.