WARNING!
Improper use of the Check O
2
Plus can cause inaccurate
oxygen readings leading to improper treatment and/or
patient harm. Follow the procedures outlined in this
user manual.
Users must become thoroughly familiar with the
information contained in this User Manual before
use. Strict adherence to the operating instructions
is necessary for safe, effective product performance.
This product will perform only as designed if operated
in accordance with the manufacturer’s operating
instructions.
DO NOT use the Check O
2
Plus to measure the
oxygen concentration of a concentrator when flowing
at rates lower than its optimal performance as
specified by the concentrator manufacturer; generally
4 LPM or less on concentrators that have a maximum
flow of 10 LPM, and 1 LPM or less on concentrators
that have a maximum flow of 5 LPM.
Use of the Check O
2
Plus near devices that generate
electrical fields may cause erratic readings.
If the Check O
2
Plus is ever exposed to liquids from
spills or immersion, immediately remove the batteries
and let the device dry completely. When dry, replace
the batteries and check for proper operation.
WARNING!
DO NOT autoclave or expose the Check O
2
Plus to
high temperatures (>60° C [140° F]).
DO NOT use ethylene oxide sterilization.
DO NOT expose the Check O
2
Plus to irradiation,
vacuum, steam, or harsh chemicals.
DO NOT expose the Check O
2
Plus to pressure
greater than 50 psi. Exposure to pressure above 50 psi
could cause leaks in the device which may adversely
affect performance in flow and pressure readings.
Repair or alteration of the Check O
2
Plus by anyone
other than a qualified technician could cause the
product to fail to perform as designed.