Invacare 212EX Oxygen Equipment User Manual


 
14VENTURE PART NO. 1093845
SPECIFICATIONS
OPERATING VOLTAGES
EXTERNAL AC/DC POWER Output: 12 + 3 VDC, 100 mA
ADAPTER: Connector: 5.5mm OD x 2.1mm
ID x 12.0mm No. 2 Connector
ALARMS Low Battery (Two [2] levels)
Cannula Disconnect (Breath not
sensed)
Device Malfunction
Low / No Gas Indicator
INDICATORS/SYMBOLS:
O
2
LED: Pulse Delivery
LED: Cannula Disconnect/Missing
Breath Indicator, Audible Alarm
LED: Battery Low Charge Indicator (1/2 to 2
hours of use time left), Audible Alarm
LED: Low or No Gas, Audible Alarm
: Class BF
BATTERY: Alkaline: (1) "D" size cell
Approx. Battery Usage Time: 80 Hours @ 20 bpm
(Variable Breath Mode)
OXYGEN DOSE DURATION:
1-variable duration breathing: Auto-adjusts between 0.5 and 1.0
seconds based on patient breathing rate.
3-fixed duration breath mode: Fixed 1.0 seconds, 0.75 seconds,
0.50 seconds
FLOWRATE RANGE (L/Min.): 0 to 6 L/Min.
OXYGEN SOURCE PRESSURE:
Venture 20: 20 + 3 psig (138 + 20 kPa)
Venture 50: 50 + 5 psig (345 + 35 kPa)
REGULATOR
Flow Settings: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,
5.0, 6.0 L/Min.
Regulated Pressure: 20 psig
O