OPERATOR ACTION DISPLAY
Turn the scale on. While “START” is displayed on the LCD (for 3 seconds) “CAL” will appear on the LCD
press the ZERO-KG/LB-KG/LB-ZERO keys one at a time on the keypad.
Press “ENTER” Enter load weight “Load/200.0”
Press “ENTER” to calibrate with 200 lb. This can be changed if desired. “CLEAR”
To change the calibration weight, press the BMI key. Then, use the arrow keys to change the value of the flashing digit.
Use the BMI key to advance to the next digit, and again use the arrow keys to change the digit to the desired value.
Repeat until you have entered the desired calibration weight. Press “ENTER”
The calibration weight must be no less than 100 lb (45 kg) and no more than 300 lb (135 kg).
Clear the weight from the platform and press “ENTER” Zero Calibration “------------”
Wait without touching or shaking the scale until the zero calibration Weight calibration “Put XXX.X”
process is finished
Load the platform with the required weight and press “ENTER” Weight calibration “------------”
Wait without touching or shaking the scale until the zero calibration process is finished. Calibration “Factor X.XXXX”
Note:
If the calibration is done in lb, the factor value you should have is
0.9000 – 1.000 If the calibration is done in kg, the factor value should be
0.4000 – 0.4250. If the factor value is different repeat the calibration procedure.
Press ENTER “DONE”/“CLEAR”/XXX
Remove weight from the platform Rebooting, please wait
Wait until the scale resumes normal operation “0.0”
6