Hanna Instruments HI 98804 Thermometer User Manual


 
1312
TO RESET PRINTING SAMPLE NUMBER
(HI98701 and HI98704 only)
This feature resets the print on demand sample number to 001.
Select the code 05
A scrolling message will be displayed.
Press the ALT and CFM keys to confirm reset or ALT and FNC to
escape.
GLP DATA
Cod. 10 - viewing GLP data
Select the code 10
A message will scroll twice across LCD.
The LCD will then display the instrument identification (ID) code.
Press # to scan remaining data, in the following order:
channel 1 last calibration date (DD.MM)
channel 1 last calibration year
Note: Data can be viewed in reverse order pressing the ! key.
Press ALT and FNC to return to function selection mode.
Code 20 - setting the identification (ID) code
When using several identical meters it may be useful to uniquely
identify them by assigning an ID code to each meter.
Select code 20. A message will scroll across LCD.
Enter a 4-digit value using the arrow keys.
Press ALT and CFM to confirm the value.
TESTING BATTERY LEVEL
Select code 41. The message "Battery test" will scroll across LCD.
If the meter is connected to an external power adapter, the LCD
will display "LINE" otherwise it will display bAtt on the upper
display, and the remaining percentage of battery charge (100%
means fully charged battery and 0% cor-
responds to the minimum battery voltage
that allows the meter to operate).
SELECTING THE THERMOCOUPLE TYPE
Three types of thermocouple can be connected to the meter:K, J and T.
Select code 70.
The types are indicated with "dEF" for K-type, "J" for J-type and
"t" for T-type. Choose the desired type using the arrow keys.
Press ALT and CFM to confirm the value.
the last printed sample number. It is then possible to select
another sample.
Press the ALT and FNC keys to return to setup mode.
Viewing logged data
Press CH/TIME (HI98804 only) or TEMP/TIME (HI98801 only) to
view data of the selected sample. Data will be displayed in the
following order:
channel 1 temperature reading
channel 2 temperature reading (HI 98804 only)
channel 3 temperature reading (HI 98804 only)
channel 4 temperature reading (HI 98804 only)
date
time
If CH/TIME or TEMP/TIME is pressed when the time is displayed,
the LCD will pass to the next Sample number.
It is then possible to scroll the data of the next sample by pressing
CH/TIME or TEMP/TIME or select a different sample by the arrow
keys.
to return to setup mode, press ALT and FNC when the meter
displays the sample number.
Cod. 02 - Lot summary printout
Select the code 02.
The message "Lot table Printing" will scroll twice across LCD.
The meter will then print a complete set of information based on
the data stored in memory:
a - lot number
b - logging interval
c - starting date
d - number of samples
e - starting time
f - logged channels numbers.
Note For lot 00 the number of samples only will be printed.
DELETE LOGGED DATA (HI98801 and HI98804 only)
Select code 05 to delete the Log on Demand data or code 06 to
delete the Timed Log Data.
A scrolling message will be displayed.
Press the ALT and CFM keys to confirm deletion.
It is also possible to escape without data deletion pressing the ALT
and FNC keys.