Graham Field JB112-070 Respiratory Product User Manual


 
JB0112-070-INS-LAB-RevB14 16
TROUBLESHOOTING
If your Neb-u-Tyke Neb-a-Doodle Nebulizer Compressor
fails to function, consult the following Troubleshooting
guide. If the problem persists, consult your Graham-Field
authorized distributor.
Problem Cause and Solution
Device doesn’t
operate
Check if plug is properly t into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Nebulization is
weak
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if lter is dirty; replace lter as needed.