8 SPECIFICATIONS
Flow Range: 1 to 5L/min
Oxygen Concentration: 1-4 LPM 94% ± 2%
5 LPM 92% ± 2%
Maximum Outlet Pressure: 6.52 psi ± .065 psi
(45kPa ± 4.5kPa)
Pressure Relief Mechanism Operational at:
36.26 psi ± 3.63 psi (250kPa ± 25kPa)
Altitude: Up to 5,997 feet (1,828 meters) above sea level
without degradation of concentration levels. Efciency
of 90% from 5,997 to 13,123 feet (1,828 to 4,000 meters)
above sea level.
Sound Level: ≤ 55dB
Power Supply: JB0160-010B, JB0160-010, JB0160-015B,
JB0160-015: AC120V±10% (60Hz) (4.6amps).
JB0160-010B-220, JB0160-010-220, JB0160-015B-220,
JB0160-015-220: AC220V±10% (50Hz) (2.2amps)
Input Power: 400 Watts Average
Weight: 57lbs (26kg)
Dimensions: H 27” x W 18” x D 15”
Minimum Warm-Up Time Before Use: 30 minutes
Electrical Classication: II Class B Type
System: Continuous
JB0160-015B, JB0160-015, JB0160-015B-220,
JB0160-015-220 only: O2 sensor (electro-chemical de-
vice): Yellow fault light illuminates when O
2
concen-
tration is greater than 71% and less than 82%; system
works. Red warning light illuminates when O
2
concentra-
tion is less than 71%; system shuts off.
Normal Operation Conditions
Temperature range: 41°F to 104°F (5°C to 40°C)
Relative humidity: 20% to 60%
Atmospheric pressure: 12.47 psi to 15.37 psi
(86kPa to 106kPa)