Graham Field 222 Blood Pressure Monitor User Manual


 
Labtron 222 / V222B Wallmax Professional Aneroid Sphygmomanometer 6
Inflating the cuff: Close the bulb’s air valve by turning the
air release valve clockwise. Squeeze the ination bulb at a
steady rate until the gauge’s needle points at approximate-
ly 30mmHg above the individual’s normal systolic pres-
sure value. If the individual’s normal blood pressure is not
known, it is recommended to inate to 200mmHg.
Systolic blood pressure reading: Open the air release
valve slowly by turning it counter-clockwise while holding
the diaphragm of the stethoscope over the brachial artery.
Proper deation rate is vital for an accurate reading. The
recommended deation rate is 2-3mmHg per second, or
a drop of one to two marks on the pressure gauge with
each heartbeat. Do not keep the cuff inflated any longer
than necessary. As the cuff begins to deate, listen care-
fully with the stethoscope. Note the reading on the gauge
as soon as a faint, rhythmic tapping or thumping sound is
heard. The rst sound is the systolic pressure reading. Al-
ways check with your health care provider to ensure read-
ings are performed correctly.
Diastolic blood pressure reading: Allow the pressure to
continue dropping at the same deation rate. Note the
reading on the gauge when the last audible thumping,
swishing, or blowing sound is heard; this is the diastolic
blood pressure reading. After a few seconds have passed
and no audible thumping, swishing, or blowing sounds are
heard, deate the cuff using the air release valve complete-
ly. Remove the cuff and stethoscope from the arm.
Record the systolic and diastolic readings. Repeat the
measurement two or more times to ensure accuracy. Only
a patient’s physician is qualied to analyze blood pressure.