Fusion F300S Light Therapy Device User Manual


 
60 F300S/F300SQ Ultraviolet Lamp System
2. Turn the Master POWER switch ON and press the LAMP ON switch.
3. The system will now operate from remote switches. Remote switch functions are
virtually the same as front panel switch functions.
Operation with Remote Controls
KEY:
S- Switch position does not matter.
Quick Restart Option (QRO)
The Quick Restart Option (QRO) provides the ability to restart the lamp at any time,
even immediately after the lamp has been switched to Standby mode. Thus, the
mandatory 18-second restart cycle time associated with routine operation has been
eliminated. The QRO may be installed at the time of the P300MQ power supply
construction or may be retrofitted into P300M power supplies in the field by trained
personnel.
Table 13. Remote Switch Functions
Remote Switch
Equivalent Panel
Control
Lamp Status
Switch Open Switch Closed
Lamp ON Lamp ON Standby Lamp ON
Lamp RESET LAMP OFF/RESET Default to lowest state LAMP OFF/RESET
Table 14. Operation with Remote Controls
Panel Controls on Master
Remote
Switches
Power Lamp Control Lamp ON Lamp Status
Off/Reset S S Off/Reset
ON LAMP OFF/RESET S LAMP OFF/RESET
ON
ON Remote RESET
Switch CLOSED
LAMP OFF/RESET
(overrides remote Lamp
ON switch)
ON
STANDBY
OPEN
STANDBY
CLOSED
ON OPEN
CLOSED Lamp ON