MEASURING TECHNIQUE
Here are some suggestions for improving the accuracy of your
temperature measurements:
When measuring higher than ambient temperatures, adjust
the connection between the probe and the surface until you
get the highest temperature reading.
When measuring lower than ambient temperatures, adjust the
connection between the probe and the surface until you get
the lowest temperature reading.
When measuring near ambient temperatures, make the
reading when the thermometer readout is most stable.
TROUBLESHOOTING
With no heat or cold applied to the probe, the measuring instrument
should display the ambient temperature. If the measuring
instrument does not read out properly, try the following:
1. Verify that the temperature-measuring instrument is designed
to be used with Type K thermocouples. It should have a yellow input
connector and / or be marked “K”.
2.
Check for an open circuit indicator on the measuring
instrument. Some temperature measuring instruments have a built-
in circuit to indicate if the connected probe is open. (All Fluke
instruments have this feature.) Refer to the measuring instrument’s
owners manual to see if this feature is available.
If you suspect a broken connection, use an ordinary ohmmeter to
check its continuity from pin to pin. The ohmmeter should read 10
ohms or less if there is continuity.
3.
Short the two input pins of the measuring instrument with a
piece of wire. If the instrument is functioning, it should indicate the
ambient temperature.
SCALE CONVERSIONS
Use the following equation to convert
°
C to
°
F:
(
°
C x 1.8) ± 32 =
°
F
Use the following equation to convert
°
F to
°
C:
(
°
F –32) x 0.5556 =
°
C