12
Wrap Thermocouple Probe (model-specific)
Type..............................................................................Type K with miniconnector and
Velcro strap, ASTM E230-03
Standard Tolerance
Measurement Range ....................................................0 °C to 100 °C (32 °F to 212 °F)
Accuracy.......................................................................±2.2 °C (4.0 °F)
Total Length..................................................................505 mm (20 in) cable terminated with
a Type K thermocouple inside a
495 mm (19.5 in) nylon Velcro cuff
Bead Thermocouple Probe (model-specific)
Type..............................................................................Type K with miniconnector
Measurement Range .................................................... -40 °C to 260 °C (-40 °F to 500 °F)
Accuracy.......................................................................±1.1 °C (2.0 °F) from 0 °C to 260 °C
(32 °F to 500 °F). Typically within
1.1 °C
(2.0 °F) from -40 °C to 0 °C
(-40 °F to 32 °F)
Cable Length ................................................................1 m (40 in) terminated with Type K
thermocouple beads
Electrical
Power Supply................................................................2 AA Batteries (alkaline or NiCD)
Power Consumption......................................................At least 12 hours battery life
Physical
Weight...........................................................................0.322 kg (0.7099 lb)
Size...............................................................................17.69 cm (6.965 in) x 16.36 cm
(6.441 in) x 5.18 cm (2.039 in)
Environmental
Operating Temperature Range.....................................0 °C to 50 °C (32 °F to 120 °F)
Relative Humidity..........................................................0 to 90 %, noncondensing up to
30 °C (86 °F)
Storage Temperature....................................................-20 °C to 65 °C (-4 °F to 150 °F)
Optional Accessories....................................................Soft Case