Fluke 189 Thermometer User Manual


 
7-1
Chapter 7
Specifications
Safety and Compliances
Maximum voltage between any terminal and earth
ground.
1000 V dc or rms ac
Compliances - DUAL RATINGS Complies with IEC 1010-1 to 1000 V Overvoltage Category III,
Pollution Degree 2; and IEC 664-1 to 600 V Overvoltage Category
IV, Pollution Degree 2 *
Certifications (Iisted and pending) CSA per standard CSA/CAN C22.2 No. 1010.1-92
UL per standard UL 3111 TÜV per standard EN 61010 Part 1-1993
Surge Protection 8 kV peak per IEC 1010.1-92
Fuse Protection for mA or µA inputs
Fuse Protection for A input
0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse
11 A, 1000 V FAST Fuse
Markings
, ,UL, TÜV and
#
* OVERVOLTAGE (Installation) Categories refer to the level of Impulse Withstand Voltage protection provided at the specified
Pollution Degree.
Overvoltage Category III equipment is equipment in fixed installations. Examples include switch gear and polyphase motors.
Overvoltage Category IV equipment is equipment for use at the origin of the installation. Examples include electricity meter
and primary over-current protection equipment.