Fluke 1502A Thermometer User Manual


 
4 Quick Start
This section briefly explains the basics of setting up and operating your 1502A
thermometer readout.
4.1 Unpacking
Unpack the thermometer carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that the following components are present:
1502A Thermometer
Extra Probe Connector
Power Cord
User’s Guide
Probe (optional—must be purchased separately)
4.2 Power
Your 1502A is configured for either 115 VAC (±10%) operation or 230 VAC
(±10%) operation. Be careful to only connect the 1502A to a mains supply of
the correct voltage. Otherwise, the instrument may be damaged. The required
voltage is indicated on the back of the 1502A. Power requirements are listed in
Section 3.1, Specifications. The IEC type power cord connects to the back of
the 1502A. The cord must be plugged in to a grounded outlet. The power
switch is located at the back of the 1502A.
When the 1502A is powered on, wait briefly while it initializes. It will then be
-
gin measuring and displaying temperature.
Because of the quality of the components used in the 1502A, it exhibits nearly
negligible drift as it warms up. The warm-up drift is typically less than 5 ppm.
Nevertheless, to ensure the best accuracy and stability, you may want to allow
the 1502A to warm up for ten minutes before use.
Accurate measurement requires that the probe be connected properly to the in
-
put and the correct probe characterization set.
4.3 Connecting the Probe
The RTD or SPRT probe connects to the back of the 1502A using a five-pin
DIN plug. Figure 1 shows how a four-wire probe is wired to the five-pin DIN
connector. One pair of wires attaches to pins 1 and 2 and the other pair attaches
11
4 Quick Start
Unpacking